Product Specification
Parameters Measured | ECG, RESP, NIBP, SPO2, Temperature, ETCO2, IBP, PR |
Suitable For | Adult, Pediatric, Neonatal |
Display | Digital |
Model | M707 |
Automation Grade | Automatic |
Size | 12.1" LED |
I Deal In | New Only |
Brand | Allied Meditec |
Product Description
Features:
Display:
ECG:
Respiration:
Respiration Rate Measuring and Alarm Range:
IBP:
- The choice of 12.1” LED color display provides superior flexibility for different clinical applications
- Accurate monitoring of ECG, SPO2, NIBP, TEMP, RESP, HR, PR and optional IBP, EtCo2
- Pacemaker detection, Electrosurgical interference proof
- Defibrillation protection and defibrillation synchronization
- OxyCRG available to judge the respiration and circular function for neonates
- 120 hours graphic and tabular trends of all parameters
- USB ports and SD card slot to ensure data storage/function upgrades
- Select from either Sidestream, or Mainstream EtCO2 technology to obtain fast, simple, and accurate CO2 results
- Nurse call function and Bi-directional communication with Central
Display:
- Various working interface selectable: Standard Monitoring Display, Large Font Intensive Care Display, Trend Graph/ Monitoring Co-Display, Bed to Bed view Display (optional), OxyCRG Dynamic View Display, Drug Dose Calculation Interface
- Sweep Speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
ECG:
- Lead Mode: 3 Leads (R, L, F or RA, LA, LL), 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
- Waveform: 7 waveforms display on a screen
- 4 selectable plus: 0.25, 0.5, 1, 2 cm/mV
- PACE detection Detection leads: 7 selectable leads: I, II, III, AVR, AVL, AVF, V
- Defibrillation energy distributary Input voltage range ±8mV, polarized voltage: ±500mV
- CMRR Diagnosis: >100 dB (no 50Hz/60Hz software wave trap)
- Monitor: >110 dB (has 50Hz/60Hz software wave trap)
- Surgery: >100dB (50Hz/60Hz software wave trap)
- System noise: <30μV p-p (RTI)
- HR Measuring and Alarm Range
- Adult/Ped: 15 bpm – 300bpm
- Neo: 15 bpm – 350bpm
- Accuracy: ±1% or ±1bpm, which great
- Resolution: 1 bpm
- Sensitivity: > 200 μV P-P
- Differential Input Impedance: ≥ 5 MΩ
- Electrode offset potential: 300mVd.c. – 600mVd.c.
- Patient leakage current: < 10 uA
- ECG Signal Range: 6 mV (Vp-p).
- ST Segment Monitoring Range Measure and Alarm: -2.0 – +2.0 mV
Respiration:
- Method: Impedance between R-F (RA-LL), R-L (RA-LA)
Respiration Rate Measuring and Alarm Range:
- Adult: 0 rpm – 120 rpm Neo/Ped 0 rpm – 150 rpm
- Resolution: 1 rpm Accuracy 2 rpm
- Gain Selection: 0.25, 0.5, 1, 2, 3, 4, 5
IBP:
- Channel: Upto 4 channel
- Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
- Static Pressure Measuring Range: (-)50 – +300 mmHg (up to 350 mmHg)
- Static Pressure Accuracy: ±2% or ±1 mmHg whichever is higher
- Dynamical Pressure Measuring Range: (-)50 – +300 mmHg
- Dynamical Pressure Accuracy: 2 % or 1mmHg which bigger
- ART: 0 – 300 (mmHg) PA 6 – 120 (mmHg)
- CVP/RAP/LAP/ICP: (-)10 – 40 (mmHg) P1/P2 (-)50 – 300 (mmHg)
- Resolution: 1 (mmHg) Zero range 200 mmHg